How is the intrinsic neutron detection efficiency of a CdZnTe detector determined using foil activation?

Blog / Date: September 11, 2025 / Author: CdZnTe.com / Hits: 31

## Determining the Intrinsic Neutron Detection Efficiency of a CdZnTe Detector Using Foil Activation


The intrinsic neutron detection efficiency of a CdZnTe (Cadmium Zinc Telluride) detector can be determined by using a technique known as foil activation. This method involves irradiating a neutron-sensitive foil with thermal neutrons and measuring the induced radioactivity in the foil. The gamma radiation emitted during the decay of the activated foil can then be detected by the CdZnTe detector. The process allows for the determination of the efficiency of the detector in capturing neutrons and converting them into measurable signals.

## 1. Understanding Foil Activation


Foil activation involves using materials (usually thin foils of neutron-capturing elements) that undergo nuclear reactions when exposed to neutrons. The most common process used in foil activation for neutron detection is the neutron capture reaction, where a neutron is absorbed by an atom, leading to the emission of gamma rays or the creation of a radioactive isotope.

For neutron detection using CdZnTe detectors, materials like Cd, B, or Li are often used as activation foils, with 113Cd being one of the most commonly employed isotopes due to its high neutron capture cross-section.

The specific reaction used in the activation process for CdZnTe detectors is:

$$
^{113}Cd + n
ightarrow ^{114}Cd^*
ightarrow ^{114}Cd + gamma
$$

This reaction produces a gamma-ray as the 114Cd nucleus de-excites. By detecting the gamma rays emitted from the activated foil, the efficiency of the CdZnTe detector in capturing neutrons can be evaluated.

## 2. Steps to Determine Neutron Detection Efficiency


The process of determining the intrinsic neutron detection efficiency using foil activation involves the following steps:

## 2.1 Preparation of the Activation Foil


A thin foil of neutron-sensitive material, such as 113Cd, is selected based on its high neutron capture cross-section. The foil is placed in close proximity to the CdZnTe detector for irradiation. The thickness and material of the foil are chosen such that the neutron flux interacting with the foil is sufficient to induce measurable radioactivity.

## 2.2 Neutron Irradiation


The foil is exposed to a known neutron flux in a neutron source, such as a nuclear reactor or a neutron generator. During this exposure, the neutrons interact with the foil, causing the activation reaction (e.g., 113Cd (n, γ) 114Cd). The neutron flux and energy spectrum should be well-characterized to ensure accurate efficiency measurements.

## 2.3 Gamma Ray Detection


After the irradiation period, the activated foil will begin to emit gamma rays as the 114Cd isotope undergoes radioactive decay. The CdZnTe detector is positioned to measure the gamma rays emitted by the activated foil. The gamma spectrum recorded by the detector will include the prompt gamma rays emitted during neutron capture, as well as any subsequent decay gamma rays.

## 2.4 Calculation of the Neutron Detection Efficiency


The intrinsic neutron detection efficiency of the CdZnTe detector is determined by comparing the measured gamma-ray intensity with the known neutron flux. Several factors need to be taken into account during the calculation:

* Gamma-ray Yield: The yield of gamma rays emitted from the foil during the activation process is proportional to the number of neutrons captured. This can be calculated based on the neutron capture cross-section of the foil material and the total number of neutrons absorbed.

* Detector Response: The gamma-ray response of the CdZnTe detector is characterized by the energy resolution and detection efficiency for gamma rays. This response can be determined through calibration with known gamma-ray sources.

* Background Subtraction: Background gamma radiation from other sources in the environment or from the neutron source itself must be subtracted from the measured gamma spectrum to isolate the signal from the activated foil.

## 2.5 Efficiency Formula


The intrinsic neutron detection efficiency η can be expressed as:

$$
η = frac{N_{gamma}}{N_{ ext{neutrons}}}
$$

Where:

* $N_{gamma}$ is the number of gamma rays detected by the CdZnTe detector (after background subtraction).
* $N_{ ext{neutrons}}$ is the number of neutrons interacting with the foil, which is related to the neutron flux and the foil's neutron capture cross-section.

This calculation provides the intrinsic efficiency of the CdZnTe detector in detecting thermal neutrons, factoring in the detector's response to gamma rays.

## 3. Advantages of Foil Activation for Neutron Detection Efficiency


* Direct Measurement: Foil activation provides a direct way to measure the neutron detection efficiency by using a well-understood activation process and gamma-ray emission.

* Calibration: It can be used as a calibration method for CdZnTe detectors, especially for space-based missions or field deployments, where precise neutron fluxes are difficult to measure directly.

* Non-Destructive: The foil activation process is non-destructive to the CdZnTe detector, making it a convenient method for evaluating detector performance without damage.

* Versatility: The technique can be applied to different materials, and by varying the material of the activation foil (such as 113Cd, B, or Li), it can be used for a wide range of neutron detection scenarios.

## 4. Challenges and Considerations


While foil activation is a powerful technique for measuring neutron detection efficiency, several factors must be carefully controlled:

* Neutron Spectrum: The efficiency measured is sensitive to the energy spectrum of the neutrons. For accurate efficiency calculation, the neutron spectrum should match the operational conditions of the detector.

* Detector Geometry: The position and geometry of both the CdZnTe detector and the activation foil need to be optimized for accurate efficiency measurements.

* Background Radiation: It is important to account for any background gamma radiation from external sources or the irradiation process, which can influence the gamma-ray spectrum and efficiency calculation.

## 5. Conclusion


Foil activation is an effective and widely used method for determining the intrinsic neutron detection efficiency of CdZnTe detectors. By measuring the gamma rays emitted from neutron capture reactions in the foil, the efficiency of the CdZnTe detector in detecting thermal neutrons can be quantified. This method provides valuable data for calibration and performance assessment in a variety of neutron detection applications, including nuclear security, environmental monitoring, and space-based missions.



CdZnTe Association (CdZnTe.com)
https://www.cdznte.com/blog/how-is-the-intrinsic-neutron-detection-efficiency-of-a-cdznte-detector-determined-using-foil-activation.html
CZT Detectors
Collection of a curated range of high-performance cadmium zinc telluride (CZT) radiation detectors for medical, security, and scientific applications
CZT Detector Bare Die for Gamma Ray
CZT Detector Bare Die for Gamma Ray
MFG: Prosun Semiconductor
Hits: 290
CZT Detector Bare Die for X Ray
CZT Detector Bare Die for X Ray
MFG: Prosun Semiconductor
Hits: 7438
CZT Substrate for HgCdTe-Based Infrared Detection
CZT Substrate for HgCdTe-Based Infrared Detection
MFG: Prosun Semiconductor
Hits: 181
GalaxRay RIID high-resolution CZT-based Personal Radiation Device
GalaxRay RIID high-resolution CZT-based Personal Radiation Device
MFG: NUVIATech Instruments
Hits: 35
NuVISION-MAX CZT-based high performance gamma imager
NuVISION-MAX CZT-based high performance gamma imager
MFG: NUVIATech Instruments
Hits: 46
NuVISION compact portable spectrometric gamma camera based on CZT
NuVISION compact portable spectrometric gamma camera based on CZT
MFG: NUVIATech Instruments
Hits: 42