What impact does leakage current have on the energy resolution of CdZnTe detectors?

Blog / Date: October 11, 2025 / Author: CdZnTe.com / Hits: 13

## Introduction to Leakage Current and Energy Resolution in CdZnTe Detectors


In semiconductor-based radiation detectors, such as those made from CdZnTe (Cadmium Zinc Telluride), the energy resolution is a critical parameter, directly linked to the detector's ability to accurately distinguish between different energy levels of incoming radiation. Leakage current, typically observed as unwanted current flowing in the reverse bias region, can severely impact the detector's performance, including its energy resolution. The energy resolution is a measure of how clearly the detector can differentiate between two closely spaced energy peaks. An increase in leakage current can degrade this resolution, reducing the detector's sensitivity and overall performance.

## How Leakage Current Affects Energy Resolution


## 1. Increased Noise and Signal-to-Noise Ratio (SNR) Reduction


* Noise Contribution: Leakage current introduces additional noise into the detector system, which interferes with the measurement of the signal produced by the radiation event. This noise is typically referred to as "dark current" in the reverse bias regime and can manifest as an unwanted signal that fluctuates over time.
* SNR Degradation: The presence of leakage current increases the baseline noise, reducing the signal-to-noise ratio (SNR) of the detector. In a radiation detector, the SNR is a crucial factor in determining the energy resolution. As the leakage current adds noise to the signal, it becomes harder to distinguish between the actual radiation-induced signal and the noise, leading to broadening of the energy peaks.
* Result: This broadening of peaks results in a reduction in energy resolution because the detector is no longer able to accurately delineate the energy of incoming radiation. The broader peaks are less distinct, and the system is less sensitive to small changes in energy.

## 2. Distortion of Pulse Shape


* Pulse Shape Distortion: In detectors like CdZnTe, each incoming photon or particle generates a charge pulse proportional to its energy. This charge pulse is then read out as a voltage signal. Leakage current can cause distortions in the pulse shape, as it contributes a constant current or noise signal that modifies the charge pulse from the radiation event.
* Impact on Pulse Processing: The distorted pulse shape can affect the way the detector electronics process the signal. Many radiation detectors use pulse shaping to extract energy information from the signal. Leakage current-induced pulse distortions can reduce the accuracy of energy measurement and lead to an improper calculation of the pulse height, further contributing to the broadening of the energy spectrum.

## 3. Reduction of Charge Collection Efficiency


* Competing with Signal Charge: Leakage current is effectively an additional flow of charge that competes with the charge generated by the radiation event. In CdZnTe detectors, which rely on the collection of charge carriers (electrons and holes) produced by ionizing radiation, leakage current can reduce the net charge collected during an event.
* Lower Charge Collection Efficiency: With leakage current drawing away charge carriers, the overall charge collection efficiency decreases. This results in a smaller signal being generated for each detected radiation event. A lower signal level combined with increased noise results in broader energy peaks and, consequently, poorer energy resolution.

## 4. Increased Spectral Broadening


* Spectral Broadening Effects: The additional current caused by leakage contributes to the overall noise level in the system, leading to spectral broadening. This effect is similar to an increase in the full width at half maximum (FWHM) of the energy peaks in the spectrum. As leakage current increases, the peaks in the energy spectrum become wider and less distinct, reducing the ability of the detector to resolve closely spaced energy levels.
* Effect on Resolution: In practice, the effect of leakage current on energy resolution is most pronounced when measuring low-energy events or weak signals. These events are more easily obscured by noise, leading to a higher FWHM value for the energy peaks and reducing the ability of the detector to distinguish between low-energy radiation events.

## 5. Thermal Effects and Drift


* Thermal Generation of Leakage Current: Leakage current often increases with temperature, as thermally generated charge carriers contribute to the overall current. In CdZnTe detectors, this can lead to an increase in leakage current at higher operating temperatures.
* Drift in Energy Spectrum: As leakage current rises, the noise level increases, leading to a drift in the energy spectrum over time. This drift further complicates energy resolution, as it may cause the energy peaks to shift, broadening them and making it harder to determine the exact energy of the radiation.

## Impact of Leakage Current on Energy Resolution in Practical Terms


The effect of leakage current on energy resolution is typically observed as a deterioration in the detector's ability to produce sharp, well-defined peaks in the energy spectrum. For example:

* Ideal Detector Resolution: In the absence of leakage current, a CdZnTe detector might be capable of achieving energy resolution on the order of 1–2% (FWHM) at a given photon energy.
* With Leakage Current: The presence of significant leakage current can increase the FWHM of the energy peaks, resulting in a much broader spectrum, which might increase the resolution to 3–5% or more, depending on the magnitude of the leakage current.

## Conclusion


Leakage current has a significant impact on the energy resolution of CdZnTe detectors by introducing noise that reduces the signal-to-noise ratio, distorting pulse shapes, decreasing charge collection efficiency, and contributing to spectral broadening. The broader peaks that result from leakage current make it more difficult for the detector to resolve energy levels accurately, leading to a reduction in the ability to discriminate between radiation events. Reducing leakage current, through better material quality, optimized crystal growth, effective passivation, and temperature control, is critical to improving the energy resolution and overall performance of CdZnTe-based radiation detectors.



CdZnTe Association (CdZnTe.com)
https://www.cdznte.com/blog/what-impact-does-leakage-current-have-on-the-energy-resolution-of-cdznte-detectors.html
CZT Detectors
Collection of a curated range of high-performance cadmium zinc telluride (CZT) radiation detectors for medical, security, and scientific applications
CZT Detector Bare Die for Gamma Ray
CZT Detector Bare Die for Gamma Ray
MFG: Prosun Semiconductor
Hits: 291
CZT Detector Bare Die for X Ray
CZT Detector Bare Die for X Ray
MFG: Prosun Semiconductor
Hits: 7441
CZT Substrate for HgCdTe-Based Infrared Detection
CZT Substrate for HgCdTe-Based Infrared Detection
MFG: Prosun Semiconductor
Hits: 181
GalaxRay RIID high-resolution CZT-based Personal Radiation Device
GalaxRay RIID high-resolution CZT-based Personal Radiation Device
MFG: NUVIATech Instruments
Hits: 35
NuVISION-MAX CZT-based high performance gamma imager
NuVISION-MAX CZT-based high performance gamma imager
MFG: NUVIATech Instruments
Hits: 46
NuVISION compact portable spectrometric gamma camera based on CZT
NuVISION compact portable spectrometric gamma camera based on CZT
MFG: NUVIATech Instruments
Hits: 42