What role do deep-level traps play in the polarization process and count rate relaxation of CdZnTe detectors?

Blog / Date: August 18, 2025 / Author: CdZnTe.com / Hits: 13

## Role of Deep-Level Traps in the Polarization Process of CdZnTe Detectors


Deep-level traps in CdZnTe detectors significantly influence the polarization process, which arises when charge carriers (electrons and holes) accumulate near the electrodes during X-ray exposure. These traps are located within the bandgap of CdZnTe and can capture charge carriers, reducing the effective charge collection efficiency and distorting the internal electric field. Here’s how deep-level traps impact polarization:

## 1. Charge Carrier Trapping and Space Charge Formation


* Trapping Mechanism: When X-ray photons interact with the CdZnTe material, they generate electron-hole pairs. In ideal conditions, these charge carriers would quickly drift toward the electrodes under the influence of the applied electric field. However, deep-level traps, often associated with defects like dislocations, vacancies, or impurity states, can capture and hold onto some of these carriers, preventing them from reaching the electrodes.

* Space Charge Buildup: As more charge carriers are trapped by deep-level defects, a space charge accumulates near the electrodes. This trapped charge creates a localized electric field that opposes the applied bias field, reducing the net electric field in the crystal and hindering the drift of subsequent charge carriers. This effect leads to polarization of the detector, which manifests as a decreased ability to collect charge carriers and an overall reduction in the detector’s response.

## 2. Distortion of the Electric Field


* Deep-level traps not only capture carriers but also contribute to the distortion of the electric field near the electrodes. The trapped charges create regions of low field strength, which interfere with the collection of charge carriers generated later during exposure. This phenomenon leads to polarization, making the detector less efficient at converting X-ray photons into measurable electrical signals.

## 3. Effect of Trap Energy Levels


* The specific role of deep-level traps depends on their energy levels within the bandgap. Trap levels located closer to the conduction band may capture electrons, while those closer to the valence band may trap holes. The type of trapped charge (electrons or holes) and the depth of the trap (how easily carriers can escape) determine how significantly polarization affects the detector’s response.

* For instance, traps with high energy levels that are deep within the bandgap have a lower probability of releasing the trapped carriers, which leads to longer polarization times and slower recovery of the detector’s performance after high-dose exposure.

## Role of Deep-Level Traps in Count Rate Relaxation


Count rate relaxation refers to the process by which a CdZnTe detector returns to its normal operating efficiency after exposure to X-rays, particularly when there is a mutation in the dose rate or intensity. Deep-level traps play a key role in this relaxation process by controlling how quickly the trapped charge carriers are released, neutralized, or redistributed. Here’s how they influence count rate relaxation:

## 1. Slow Carrier Release


* Detrapping Process: After an X-ray dose mutation or intense exposure, trapped carriers need to be released or neutralized to restore the detector’s normal performance. Deep-level traps have varying thermal activation energies, meaning some traps can release carriers more easily than others, depending on temperature and electric field conditions. Traps that are deeper in the bandgap tend to release carriers more slowly, which prolongs the relaxation time and delays the recovery of the detector’s count rate.

* Thermal Activation of Traps: At higher temperatures, the thermal energy can assist in releasing the trapped carriers by providing them with enough energy to overcome the trap’s potential barrier. The rate of detrapping increases with temperature, which accelerates the recovery process and reduces the relaxation time. However, at low temperatures, deep traps may retain carriers for longer periods, causing a slower recovery.

## 2. Effect of Trap Density


* The density of deep-level traps also influences the relaxation time. A high concentration of traps leads to a larger accumulation of trapped carriers during X-ray exposure, which in turn requires more time to neutralize or release these carriers. This causes a prolonged relaxation time. Therefore, the presence of a high density of deep-level traps results in a slower recovery of the detector’s count rate after high-dose exposures.

## 3. Space Charge Neutralization


* As the trapped carriers are gradually released or recombine, the space charge near the electrodes is neutralized, allowing the internal electric field to return to its normal state. This neutralization process is essential for the detector to regain its full charge collection efficiency. The speed at which space charge is neutralized depends on the deep-level traps’ ability to release carriers, and this is influenced by the nature of the traps (e.g., their energy levels and density).

## 4. Impact on Detector Performance During High Dose Rates


* Increased Polarization at High Dose Rates: Under high-dose X-ray exposure, the polarization effect becomes more pronounced because more charge carriers are generated and trapped. The greater the number of trapped carriers, the greater the distortion in the internal electric field. As a result, detectors with a higher concentration of deep-level traps may exhibit more significant polarization effects and longer relaxation times, leading to reduced efficiency and slower recovery in high-intensity environments.

## 5. Role of Bias Voltage in Relaxation


* The bias voltage applied across the CdZnTe detector can influence the release of trapped carriers. A higher bias voltage creates a stronger electric field, which can help release the trapped charge carriers more quickly and reduce the polarization effects. This accelerates the relaxation process by aiding in the movement of carriers away from the traps and towards the electrodes. In contrast, lower bias voltages may result in slower carrier drift and prolonged trapping effects.

## Conclusion


Deep-level traps in CdZnTe detectors significantly influence both the polarization process and count rate relaxation by trapping charge carriers and distorting the internal electric field. These traps create space charges that reduce the detector’s charge collection efficiency, leading to a decreased response to X-ray exposure. During count rate relaxation, the release or neutralization of trapped carriers determines how quickly the detector recovers to its normal operating state. The temperature, bias voltage, and trap density all play critical roles in modulating the behavior of these deep-level traps and, consequently, the polarization and relaxation processes. Proper management of these factors is essential for improving the performance and reliability of CdZnTe detectors, particularly in high-dose or high-flux radiation environments.



CdZnTe Association (CdZnTe.com)
https://www.cdznte.com/blog/what-role-do-deep-level-traps-play-in-the-polarization-process-and-count-rate-relaxation-of-cdznte-detectors.html
CZT Detectors
Collection of a curated range of high-performance cadmium zinc telluride (CZT) radiation detectors for medical, security, and scientific applications
CZT Detector Bare Die for Gamma Ray
CZT Detector Bare Die for Gamma Ray
MFG: Prosun Semiconductor
Hits: 290
CZT Detector Bare Die for X Ray
CZT Detector Bare Die for X Ray
MFG: Prosun Semiconductor
Hits: 7438
CZT Substrate for HgCdTe-Based Infrared Detection
CZT Substrate for HgCdTe-Based Infrared Detection
MFG: Prosun Semiconductor
Hits: 181
GalaxRay RIID high-resolution CZT-based Personal Radiation Device
GalaxRay RIID high-resolution CZT-based Personal Radiation Device
MFG: NUVIATech Instruments
Hits: 35
NuVISION-MAX CZT-based high performance gamma imager
NuVISION-MAX CZT-based high performance gamma imager
MFG: NUVIATech Instruments
Hits: 46
NuVISION compact portable spectrometric gamma camera based on CZT
NuVISION compact portable spectrometric gamma camera based on CZT
MFG: NUVIATech Instruments
Hits: 42